SAD/MAD experiments
Start program for fluorescence detector
- Open a window on pc5080
- type: Rontec
Take a fluorescence spectrum
- Choose your element and move with the "+20eV" button to the desired energy.
- Press "Take Data" button to record a first fluorescence spectrum.
- Check whether the peaks are at the expected positions.
- Use the filters in order to prevent saturation.
- The dead time percentage should read well below 0.20.
Perform the fluorescence scan
- Autochooch will suggest the energies for the peak and inflection point.
It is recommendend start the SAD/MAD with the peak wavelength,
which has to be set in the IDL window with the command sete, value .
- Return to data acquisition software and start the data collection.
- For the high-energy remote increase the energy by 100 -- 1000 eV.
- Inverse-beam data collection.