SAD/MAD experiments


Start program for fluorescence detector

  • Open a window on pc5080

  • type: Rontec

Take a fluorescence spectrum

  • Choose your element and move with the "+20eV" button to the desired energy.

  • Press "Take Data" button to record a first fluorescence spectrum.

  • Check whether the peaks are at the expected positions.

  • Use the filters in order to prevent saturation.

  • The dead time percentage should read well below 0.20.

Perform the fluorescence scan

  • Autochooch will suggest the energies for the peak and inflection point.
    It is recommendend start the SAD/MAD with the peak wavelength,
    which has to be set in the IDL window with the command sete, value .

  • Return to data acquisition software and start the data collection.

  • For the high-energy remote increase the energy by 100 -- 1000 eV.

  • Inverse-beam data collection.